FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 1961498
·
Received January 10, 2011
Report
- Report Number
- 3002158293-2011-00020
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- November 26, 2010
- Report Date
- January 7, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY PACK SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY/CHARGER FAULT) HAS BEEN CONFIRMED. THE CAUSE FOR THE BATTERY FAULT (0F80) WAS A BROKEN WHITE WIRE WITHIN THE BATTERY PACK WHERE THE WIRE ATTACHES TO THE BATTERY CELLS. THE ROOT CAUSE OF THE BROKEN WIRE HAS NOT BEEN DETERMINED, BUT MAY HAVE BEEN CAUSED BY A SEVERE SHOCK FROM DROPPING THE PACK. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK.
Description of Event or Problem · 1
A (B)(4) OLD FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT ONE OF HER BATTERIES WILL NOT FULLY CHARGE AFTER BEING ON THE CHARGER FOR 24 HOURS. THE PT WAS PROVIDED WITH A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |