FDA Adverse Event Injury Summary report: N

GORE EXCLUDER AAA ENDOPROSTHESIS

MDR report key: 1961395 · Received January 10, 2011

Report

Report Number
2017233-2011-00010
Event Type
Injury
Date Received
January 10, 2011
Date of Event
December 16, 2010
Report Date
January 10, 2011
Manufacturer
W.L. GORE & ASSOCIATES
Product Code
MIH
PMA / PMN Number
P020004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO GORE, AND AN EXPLANT EVAL IS BEING CONDUCTED. A REVIEW OF THE MFG PAPERWORK HAS BEEN CONDUCTED. THE REVIEW OF THE MFG PAPERWORK VERIFIED THAT THIS LOT MET ALL PRE-RELEASE SPECS. FURTHER INVESTIGATION IS IN PROCESS. ADD'L GORE EXCLUDER AAA DEVICES RELATED TO THIS EVENT: PXC181000/(B)(4), PXC181400/(B)(4), PXC161000/(B)(4).

Description of Event or Problem · 1

ON (B)(6), 2010, THE PT WAS IMPLANTED WITH FOUR GORE EXCLUDER AAA ENDOPROSTHESES TO TREAT AN ABDOMINAL AORTIC ANEURYSM. THE PT'S AORTIC DIAMETERS WERE WITHIN GUIDELINES OF THE GORE EXCLUDER AAA ENDOPROSTHESIS INSTRUCTIONS FOR USE, HOWEVER, THERE WAS A LOT OF THROMBUS IN THE ANEURYSM SAC, AND THE AORTA AND ILIAC ARTERIES WERE TORTUOUS. THE GORE EXCLUDER AAA TRUNK-IPSILATERAL LEG ENDOPROSTHESIS WAS PLACED WITHOUT INCIDENT, FOLLOWED BY TWO GORE EXCLUDER AAA CONTRALATERAL LEG ENDOPROSTHESES. UPON BALLOONING THE PROXIMAL PORTION OF THE TRUNK-IPSILATERAL LEG USING A GORE TRI-LOBE BALLOON, THE CONTRALATERAL LEG SLIPPED OUTSIDE THE CONTRA-GATE BY ABOUT 1MM. AN ADD'L CONTRALATERAL LEG WAS PLACED TO BRIDGE THE RESULTANT GAP. THE FINAL ANGIOGRAPHIC RUN INDICATED THAT THE TRUNK-IPSILATERAL LEG HAD MIGRATED 1.5-3CM PROXIMALLY AND COVERED THE RENAL ARTERIES. ALTHOUGH THE RENAL ARTERIES APPEARED PATENT, THE DECISION WAS MADE TO CONVERT TO OPEN SURGICAL REPAIR. ALL FOUR DEVICES WERE EXPLANTED. THE PT TOLERATED THE PROCEDURE WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GORE EXCLUDER AAA ENDOPROSTHESIS MIH, SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W.L. GORE & ASSOCIATES WLG425 8040617

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R