VISIAN ICL (IMPLANTABLE COLLAMER LENS)
Report
- Report Number
- 2023826-2011-00024
- Event Type
- Injury
- Date Received
- January 7, 2011
- Date of Event
- November 29, 2010
- Report Date
- December 15, 2010
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- MTA
- PMA / PMN Number
- P030016
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION METHOD: LENS WORK ORDER SEARCH, MEDICAL REVIEW. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. PER MEDICAL REVIEW - ACCORDING TO USE FMEA (FAILURE MODES AND EFFECT ANALYSIS) IT HAS BEEN DETERMINED THAT EXCESSIVE VAULTING IS A CONSEQUENCE OF A WRONG LENS USE FAILURE MODE (I.E. IMPROPER WHITE TO WHITE MEASUREMENT, VARIABILITY OF THE WHITE TO WHITE MEASUREMENTS BASED UPON DIFFERENT TECHNIQUES UTILIZED, IMPROPER SULCUS TO SULCUS MEASUREMENT (IF UBM USED), AND PATIENT CONDITION; POOR CORRELATION OF WHITE TO WHITE MEASUREMENT AND LENGTH OF CILIARY SULCUS IN AN INDIVIDUAL CASE; IRREGULAR CILIARY SULCUS OR CILIARY SULCUS CYST). A HIGHER THAN IDEAL VAULTING IN THE ABSENCE OF SIGN AND/OR SYMPTOMS DOES NOT NEED AN EXCHANGE. HOWEVER, THE SURGEON MAY DECIDE TO EXCHANGE IF HE DETERMINES THAT THIS CONDITION MAY AFFECT THE OUTCOME OF THE PATIENT'S VISION. (B)(4). LENS NOT RETURNED.
(B)(4) SECONDARY SURGERY. -(B)(4) EXCESSIVE. (B)(4).
THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.2MM VICM 13.2 IMPLANTABLE COLLAMER LENS IN THE PT'S LEFT EYE (OS) ON (B)(6) 2010. THE LENS WAS EXPLANTED ON (B)(6) 2010 DUE TO EXCESSIVE VAULTING. SUBSEQUENTLY, THE PT EXPERIENCED NARROWING OF THE ANGLE AND SHALLOWING OF THE ANTERIOR CHAMBER. THE ICL WAS EXCHANGED FOR A SHORTER LENS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VISIAN ICL (IMPLANTABLE COLLAMER LENS) | INTRAOCULAR LENS | MTA | STAAR SURGICAL CO. | VICM 13.2 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Required Intervention | CARTRIDGE: MODEL: UNK, LOT NUMBER: UNK| FOAM TIP PLUNGER: MODEL UNK, LOT NUMBER: UNK| INJECTOR: MODEL UNK, LOT# UNK |