FDA Adverse Event Malfunction Summary report: N

CENTRA BED

MDR report key: 1961309 · Received December 20, 2010

Report

Report Number
1824206-2010-11782
Event Type
Malfunction
Date Received
December 20, 2010
Date of Event
November 24, 2010
Report Date
November 24, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE WORN BRAKE BUSHING AND ALL OF THE CASTERS TO REPAIR THE BED.

Description of Event or Problem · 1

BRAKE WOULD NOT HOLD AFTER THE BRAKE WAS SET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTRA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 850

Patients

Seq Age Sex Outcome Treatment
1