FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 1961250 · Received January 11, 2011

Report

Report Number
9617766-2011-00058
Event Type
Malfunction
Date Received
January 11, 2011
Date of Event
February 5, 2008
Report Date
January 11, 2011
Manufacturer
GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TC
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED THE 24V POWER SUPPLY. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE C-ARM WILL NOT MOVE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLUOROSCOPIC X-RAY JAA GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD 8800

Patients

Seq Age Sex Outcome Treatment
1