FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM

MDR report key: 19611998 · Received June 26, 2024

Report

Report Number
0002023141-2024-02129
Event Type
Injury
Date Received
June 26, 2024
Date of Event
April 15, 2024
Report Date
November 12, 2024
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019508
PMA / PMN Number
K061410
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). D10. CONCOMITANT MEDICAL PRODUCTS TSVH8, IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 8MM LOT 1219851. G4: PREMARKET IDENTIFICATION K011028/K013227.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER CMP (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B3. DATE OF EVENT B4: DATE OF THIS REPORT. B5: DESCRIBE EVENT OR PROBLEM. D4: ADDITIONAL DEVICE INFORMATION. D9: DEVICE AVAILABILITY . G3: DATE RECEIVED BY MANUFACTURER. G6: TYPE OF REPORT. H1: TYPE OF REPORTABLE EVENT. H2: FOLLOW UP TYPE. H3: DEVICE EVALUATED BY MANUFACTURER. H4: DEVICE MANUFACTURER DATE. H6: ADVERSE EVENT PROBLEM. H10: ADDITIONAL NARRATIVE. ZIMVIE RECEIVED ONE (1) TSVB10, (IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE. IMPLANT FRACTURE IDENTIFIED AT THE COLLAR. PRODUCT SHOWS EXTENSIVE DAMAGE, POSSIBLY CAUSED DURING REMOVAL. BROKEN SCREW FOUND INSIDE. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1230112. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1230112 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: FRACTURE IMPLANT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE PARAFUNCTIONAL HABITS/PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

THE DOCTOR REPORTS THAT THE DENTAL IMPLANTS LOCATED IN POSITIONS NUMBER 45 AND 46 FAILED BECAUSE THEY FRACTURED BY THE HEAD. THE DOCTOR REPORTS THAT THE PROCEDURE WAS NOT CONCLUDED BY PLACING ANOTHER IMPLANTS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678442 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL 1230112 00889024019508

Patients

Seq Age Sex Outcome Treatment
1 66 YR Male DENTAL IMPLANT SEE H10 NARRATIVE| DENTAL IMPLANT SEE H10 NARRATIVE