MDR report key: 1961161 · Received January 11, 2011

Report

Report Number
9617766-2011-00046
Event Type
Malfunction
Date Received
January 11, 2011
Date of Event
November 9, 2010
Report Date
January 11, 2011
Manufacturer
GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE FUSE WAS RESET DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE STENOSCOP SYSTEM WOULD NOT TURN ON. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STENOSCOP FLUOROSCOPIC X-RAY JAA GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD STENOSCOP