FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1961158 · Received January 11, 2011

Report

Report Number
1720753-2011-00163
Event Type
Malfunction
Date Received
January 11, 2011
Date of Event
December 1, 2010
Report Date
January 11, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CANCELED THE SERVICE CALL.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM WOULD NOT BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1