FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 1961140 · Received January 11, 2011

Report

Report Number
1720753-2011-00143
Event Type
Malfunction
Date Received
January 11, 2011
Date of Event
November 30, 2010
Report Date
January 11, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CENTRAL PROCESSOR UNIT WAS REPLACED AND SOFTWARE WAS RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO BOOT UP. NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6800 FLUOROSCOPIC XRAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1