RMONIC ACE 36CM W ERG HANDLE
Report
- Report Number
- 3005075853-2011-00179
- Event Type
- Malfunction
- Date Received
- January 14, 2011
- Date of Event
- December 27, 2010
- Report Date
- January 5, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED WITH THE TISSUE PAD MELTED AND PARTIALLY DETACHED WITH THE BLADE GOUGED AT THE TIP. THE DEVICE WAS TESTED WITH A TEST HANDPIECE AND GENERATOR AND ERROR CODE 5 WAS RECEIVED, NO FURTHER FUNCTIONAL TESTING COULD BE PERFORMED. PROBABLE CAUSES OF BLADE DAMAGE ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. PROBABLE CAUSES OF THE TISSUE PAD DAMAGE IS APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. AND ACTIVATING THE BLADE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. BOTH CONDITIONS CAN RESULT IN PROBABLE DAMAGE TO THE INSTRUMENT.
(B)(4).
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMOIDECTOMY PROCEDURE, THE TISSUE PAD WAS DETACHED OFF OUTSIDE THE PATIENT. NO PIECES FELL INTO THE PATIENT. AN ERROR 5 WAS DISPLAYED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RMONIC ACE 36CM W ERG HANDLE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |