FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1960858 · Received December 17, 2010

Report

Report Number
2027969-2010-02204
Event Type
Malfunction
Date Received
December 17, 2010
Date of Event
November 22, 2010
Report Date
December 17, 2010
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
021923
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INRATIO PRECISION DATA PROVIDED BY END-USER: 1ST INR: 1.3, 2ND INR: 1.9, MEAN: 1.60, SD: 0.42, %CV: 26.52. ANALYSIS OF CUSTOMER'S DATA REVEALED THAT REPEATED INRATIO TEST RESULT COMPARISON DID NOT MEET PRECISION CRITERIA. CUSTOMER'S RESULTS ARE CONSIDERED INACCURATE WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO PRODUCT IS EXPECTED TO BE RETURNED. RETAINED STRIP TESTING FROM A PREVIOUS CASE REVEALED THAT TEST RESULT COMPARISONS MET PRECISION CRITERIA. INVESTIGATION RESULTS FROM A PREVIOUS CASE: DONOR: 1, RETURN1: 3.4, IN-HOUSE2: 3.7, IN-HOUSE3: 3.4, MEAN: 3.50, SD: 0.17, %CV: 4.95. DONOR: 2, 2.1, 2.3, 2.2, 2.20, 0.10, 4.55. FOR EACH PAIR OF REPLICATES, FOR EACH SAMPLE OF STRIP LOT 237432, MEAN AND STANDARD DEVIATIONS WERE CALCULATED. IN-HOUSE TEST RESULTS WERE 4.95% AND 4.55%, RESPECTIVELY FOR EACH DONOR AND ARE LESS THAN 16%. BOTH SAMPLES PASS THE CRITERIA FOR PRECISION. NO DISCREPANT RESULTS WERE ESTABLISHED ON RETURNED AND IN-HOUSE METERS. NO FURTHER ACTION IS REQUIRED. (B)(4). DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (B)(4), NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CALLER ALLEGED IMPRECISION WITH INRATIO METER. RESULTS AS FOLLOWS: DATE: (B)(6) 2010, INR: 1.3, 1.9. PT'S THERAPEUTIC RANGE: 2-3 INR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100139 237432

Patients

Seq Age Sex Outcome Treatment
1 NI