FDA Adverse Event Death Summary report: N

ENTRUST AT

MDR report key: 1960776 · Received January 14, 2011

Report

Report Number
6000144-2011-00007
Event Type
Death
Date Received
January 14, 2011
Date of Event
November 12, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S46
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE INITIAL REPORTED EVENT WAS RECEIVED ON 11/4/2010. OF NOTE, THE REPORTABLE MALFUNCTION IS NORMALLY SUBMITTED VIA A BIMONTHLY MEDWATCH REPORT SUBMISSION THAT WOULD HAVE BEEN DUE ON 2/10/2011 . INFORMATION WAS SUBSEQUENTLY RECEIVED ON 1/13/2011 AND REVEALED PATIENT DIED. AS THERE IS NEW INFORMATION THAT REASONABLY SUGGESTS THE DEVICE HAS OR MAY HAVE CAUSED OR CONTRIBUTED TO A DEATH, THIS EVENT NO LONGER QUALIFIES FOR BIMONTHLY REPORTING AND IS THEREFORE BEING SUBMITTED AS A 30-DAY REPORT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE INITIAL REPORTED EVENT WAS RECEIVED ON (B)(6) 2010. OF NOTE, THE REPORTABLE MALFUNCTION IS NORMALLY SUBMITTED VIA A BIMONTHLY MEDWATCH REPORT SUBMISSION THAT WOULD HAVE BEEN DUE ON (B)(6) 2011 . INFORMATION WAS SUBSEQUENTLY RECEIVED ON (B)(6) 2011 AND REVEALED PATIENT DIED. AS THERE IS NEW INFORMATION THAT REASONABLY SUGGESTS THE DEVICE HAS OR MAY HAVE CAUSED OR CONTRIBUTED TO A DEATH, THIS EVENT NO LONGER QUALIFIES FOR BIMONTHLY REPORTING AND IS THEREFORE BEING SUBMITTED AS A 30-DAY REPORT. EVALUATION SUMMARY (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. RESISTANCE/IMPEDANCE INCREASE - WEEKLY MEASUREMENT DATA SHOWS A GRADUAL INCREASE FOR MIN AND MAX RV PACE=656 TO 1024 OHMS PEAK BETWEEN (B)(6) 2010. PATIENT ALERT FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(6) 2010 02:15:03.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RV (RIGHT VENTRICULAR) PACING IMPEDANCE HAD STEADILY INCREASED OVER THE PREVIOUS 9 MONTHS. LATER REVIEW OF MANUFACTURER'S DATABASE REVEALED THE PATIENT HAD DIED. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED. FOLLOW UP WITH CLINIC REVEALED THE PATIENT WENT TO HOSPICE, THE DEVICE THERAPIES WERE TURNED OFF, AND THE PATIENT DIED THAT SAME DAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RV (RIGHT VENTRICULAR) PACING IMPEDANCE HAD STEADILY INCREASED OVER THE PREVIOUS 9 MONTHS. LATER REVIEW OF MANUFACTURER'S DATABASE REVEALED THE PATIENT HAD DIED. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTRUST AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154ATG ASKU

Patients

Seq Age Sex Outcome Treatment
1 80 YR Death