FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 1960748
·
Received December 16, 2010
Report
- Report Number
- 1811755-2010-02056
- Event Type
- Malfunction
- Date Received
- December 16, 2010
- Date of Event
- November 29, 2010
- Report Date
- November 29, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH THE MOTOR, ROTOR, DRIVESHAFT, AND BEARINGS, WHICH WERE EACH REPLACED ALONG WITH OTHER COMPONENTS. SERVICE DID A CLEAN, LUBE, AND ADJUST TO THE DEVICE, AND IT WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE READ AS OVERHEATING ON THE CONSOLE WHILE BEING TESTED. THIS DID NOT OCCUR DURING A SURGICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT OR ANY ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |