FDA Adverse Event Malfunction Summary report: N

SAGITTAL SAW ATTACHMENT

MDR report key: 1960739 · Received December 16, 2010

Report

Report Number
1811755-2010-02041
Event Type
Malfunction
Date Received
December 16, 2010
Date of Event
November 17, 2010
Report Date
November 18, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HRX
PMA / PMN Number
K943323
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER COMMENT WAS CONFIRMED. THE DEVICE HAD NUMEROUS DENTS AND MARKINGS CONSISTENT WITH ROUGH HANDLING. IF THE BLADE MOUNT BECOMES DAMAGED TO THE POINT OF BENDING OR DEFORMING, THEN BLADE WILL BIND AND CAUSE A SITUATION OF ADDED FRICTION HEAT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A ROUTINE MAINTENANCE VISIT AT A CUSTOMER, THE FIELD SERVICE TECH INDICATED THAT THE HANDPIECE GETS HOT AFTER 40 SECONDS RUN TIME. THERE WAS NO PATIENT INVOLVEMENT OR ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SAGITTAL SAW ATTACHMENT SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTACHM HRX STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK