FDA Adverse Event Injury Summary report: N

CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS

MDR report key: 1960632 · Received January 13, 2011

Report

Report Number
2015691-2011-14673
Event Type
Injury
Date Received
January 13, 2011
Date of Event
September 29, 2010
Report Date
December 14, 2010
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
PMA / PMN Number
P860057/S001
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE NOT RETURNED. ADDITIONAL MANUFACTURER NARRATIVE: UNFORTUNATELY, THE VALVE WAS NOT RETURNED TO EDWARDS FOR ANALYSIS. THE DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED AND THIS DEVICE PASSED ALL MANUFACTURING AND STERILIZATION INSPECTIONS WITH NO NONCONFORMANCE. CALCIFICATION IS A WELL RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION.

Description of Event or Problem · 1

EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE. IN THIS CASE, IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AFTER AN IMPLANT DURATION OF APPROXIMATELY 11 YEARS AND 9 MONTHS. THROUGH FOLLOW-UP WITH THE HEALTH-CARE PROVIDER, IT WAS LEARNED THAT THE DEVICE WAS EXPLANTED DUE TO AORTIC STENOSIS. ACCORDING TO THE OPERATIVE REPORT, A TRANSESOPHAGEAL ECHOCARDIOGRAM WAS PERFORMED PRE AND POST BYPASS. PRE-BYPASS THERE WAS CONFIRMATION OF A SEVERE AORTIC STENOSIS AND TRACE TO MILD MITRAL REGURGITATION. ON GROSS INSPECTION, THE PROSTHETIC AORTIC VALVE WAS HEAVILY CALCIFIED. THE LEAFLETS WERE FIBROTIC WITH RESTRICTED MOBILITY. THE VALVE WAS SHARPLY RESECTED AND REPLACED WITH A 25MM CE MAGNA BOVINE PERICARDIAL PROSTHESIS. A COMPLETE RIGHT HEART AND LEFT HEART MAZE PROCEDURE WAS PERFORMED. THE LEFT ATRIAL APPENDAGE WAS AMPUTATED AS WELL. THE PATIENT TOLERATED THE PROCEDURE WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES 2800 8J1079

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention