CLEARCUT KNIVES
Report
- Report Number
- 2523835-2024-01541
- Event Type
- Malfunction
- Date Received
- June 25, 2024
- Date of Event
- May 27, 2024
- Report Date
- September 22, 2024
- Manufacturer
- ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
- Product Code
- HNN
- UDI-DI
- 00000000000000
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION IS PROVIDED IN SECTIONS D.9, H.3, H.6 AND H.11. ONE OPENED KNIFE, WITH NO TIP PROTECTOR WAS RECEIVED FOR THE REPORT OF BLUNT. SAMPLE WAS VISUALLY INSPECTED AND FOUND TO BE NONCONFORMING, BLADE WAS OBSERVED TO HAVE A BENT TIP. PENETRATION TESTING COULD NOT BE PERFORMED DUE TO THE DAMAGE OF THE SAMPLE. A REVIEW OF THE DEVICE HISTORY RECORD TRACEABLE TO THE REPORTED LOT NUMBERS, INDICATES THAT THE REPORTED PRODUCT WAS PROCESSED AND RELEASED ACCORDING TO THE REPORTED PRODUCT¿S ACCEPTANCE CRITERIA THE EXACT ROOT CAUSE COULD NOT BE DETERMINED FROM THE INVESTIGATION PERFORMED. THE DAMAGE TO THE RETURNED SAMPLE IS CONSISTENT WITH DAMAGE THAT CAN OCCUR WHEN THE BLADE CONTACTS A HARD SURFACE SUCH AS THE PROTECTIVE BLADE TRAY WHEN PRODUCT IS IMPROPERLY REMOVED OR INSERTED AFTER USE, OR IMPROPER HANDLING, OR CONTACT WITH ANOTHER INSTRUMENT DURING SURGERY OR SET-UP. THE DAMAGE SEEN ON THE RETURNED COMPLAINT SAMPLE COULD HAVE CONTRIBUTED TO THE CUSTOMERS REPORTED ISSUE OF BLUNT. THE EXACT ROOT CAUSE FOR THE DAMAGED KNIFE SAMPLE IS UNKNOWN, THEREFORE SPECIFIC ACTION WITH REGARDS TO THIS COMPLAINT CANNOT BE TAKEN. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MANUFACTURING. ANY NONCONFORMANCE, SUCH AS THE DAMAGED TIP EXHIBITED ON THE RETURNED OPENED SAMPLE IS REMOVED FROM THE LOT AND SCRAPPED. FUNCTIONAL PENETRATION TESTING IS PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. COMPLAINTS ARE REVIEWED AND MONITORED AT REGULAR INTERVALS FOR ANY SIGNIFICANT ADVERSE TRENDS. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS:(B)(4) H.10 REFLECTS ALL RELATED REPORT NUMBERS ASSOCIATED WITH THIS PRODUCT EVENT THAT HAVE BEEN SUBMITTED AT THIS TIME.
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING CATARACT SURGERY ON THE LEFT EYE, AN OPHTHALMIC KNIFE WAS BLUNT AND NOTICED IMMEDIATELY WHEN THE INCISION WAS MADE. SURGERY WAS COMPLETED ON THE SAME DAY WITH NO PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 697806 | CLEARCUT KNIVES | KNIFE, OPHTHALMIC | HNN | ALCON RESEARCH, LLC - ALCON PRECISION DEVICE | NA | 166E90 | 00000000000000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | CUSTOM-PAK SURGICAL PROCEDURE PACK |