8MM MCS TIP COVER ACCESSORY
Report
- Report Number
- 2955842-2011-00013
- Event Type
- Malfunction
- Date Received
- January 13, 2011
- Date of Event
- December 16, 2010
- Report Date
- December 16, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA
- Reporter Occupation
- NURSE
Narratives
AN INVESTIGATION CONDUCTED BY THE FIELD SERVICE ENGINEER (FSE) WAS UNABLE TO REPLICATE THE CUSTOMER REPORTED ARCING PROBLEM. THE DA VINCI SYSTEM WAS FOUND TO BE WITHIN SPECIFICATION AND THE ESU SETTINGS USED DURING THE PROCEDURE WERE WITHIN THE ISI RECOMMENDED RANGE. DIFFERENT MCS INSTRUMENTS AND TIP COVER ACCESSORIES WERE USED DURING THE PROCEDURE AND THESE ITEMS WERE UNAVAILABLE DURING FSE'S SITE VISIT. THE TIP COVER ACCESSORIES HAVE BEEN DISCARDED BY THE SITE AND MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENTS ARE NOT AVAILABLE FOR RETURN, THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SUBMITTED. AS OF JANUARY 13, 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL. THE FOLLOWING MEDWATCH MFR REPORTS WERE SUBMITTED TO THE FDA REGARDING THESE EVENTS: 2955842-2011-00012 AND 2955842-2011-00014.
IT WAS REPORTED THAT DURING A DA VINCI S SURGICAL PROCEDURE, WHILE USING THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WITH THE TIP COVER ACCESSORY INSTALLED, ARCING WAS OBSERVED. THE MCS INSTRUMENT AND TIP COVER WERE REMOVED AND REPLACED, AND THE PROCEDURE CONTINUED AS PLANNED. AFTER AN UNSPECIFIED AMOUNT OF TIME, THE REPLACEMENT MCS INSTRUMENT WITH THE TIP COVER ACCESSORY INSTALLED WAS OBSERVED TO HAVE ARCED. THE MCS INSTRUMENT AND TIP COVER WERE REMOVED, THE MCS INSTRUMENT WAS INSPECTED AND FOUND TO HAVE NO DAMAGE AND THEREFORE ONLY THE TIP COVER WAS REPLACED. THE SURGICAL PROCEDURE RESUMED AS PLANNED. A SHORT TIME LATER, THE SURGICAL STAFF WITNESSED ARCING FROM THE MCS INSTRUMENT WITH THE TIP COVER ACCESSORY INSTALLED. THE MCS INSTRUMENT AND TIP COVER WERE REMOVED A THIRD TIME AND THE TIP COVER WAS REPLACED. THE PROCEDURE WAS COMPLETED WITH THE FOURTH REPLACEMENT TIP COVER ACCESSORY AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 8MM MCS TIP COVER ACCESSORY | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY | NAY | INTUITIVE SURGICAL,INC. | 400180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI S SYSTEM, INSTRUMENTS, ACCESSORIES, & ES |