MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP EAST EUROPE
Report
- Report Number
- 1423500-2011-00569
- Event Type
- Malfunction
- Date Received
- January 13, 2011
- Date of Event
- November 16, 2010
- Report Date
- December 17, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HU
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THIS REPORT FOR A TRANSFER SET LEAK WAS CONFIRMED IN THE LAB. THE RESULTS OF A SAMPLE EVALUATION REVEALED THAT ONE OF THE OCCLUDER FEET WAS BROKEN. DURING VISUAL INSPECTION, IT WAS NOTED THAT THERE WAS NO WHITISH SPOT ON THE OCCLUDER LEG THAT IS INDICATING A POSSIBLE OVERTORQUING. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A ROOT CAUSE WAS NOT IDENTIFIED DUE TO INSUFFICIENT INFORMATION. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, HOWEVER THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED. A 510K NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510K NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS THE SAME OR SIMILAR TO THE PRODUCT DISTRIBUTED WITHIN THE U.S.
A PATIENT REPORTED TO BAXTER-(B)(4) THAT ON (B)(6) 2010 THE TRANSFER SET LEAKED. ACCORDING TO THE PATIENT, THE TWIST CLAMP WAS DAMAGED. THE TRANSFER SET WAS REPLACED ON (B)(6) 2010. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP EAST EUROPE | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME | H09I16081 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |