FDA Adverse Event Malfunction Summary report: N

MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP EAST EUROPE

MDR report key: 1960547 · Received January 13, 2011

Report

Report Number
1423500-2011-00569
Event Type
Malfunction
Date Received
January 13, 2011
Date of Event
November 16, 2010
Report Date
December 17, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THIS REPORT FOR A TRANSFER SET LEAK WAS CONFIRMED IN THE LAB. THE RESULTS OF A SAMPLE EVALUATION REVEALED THAT ONE OF THE OCCLUDER FEET WAS BROKEN. DURING VISUAL INSPECTION, IT WAS NOTED THAT THERE WAS NO WHITISH SPOT ON THE OCCLUDER LEG THAT IS INDICATING A POSSIBLE OVERTORQUING. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A ROOT CAUSE WAS NOT IDENTIFIED DUE TO INSUFFICIENT INFORMATION. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, HOWEVER THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED. A 510K NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510K NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS THE SAME OR SIMILAR TO THE PRODUCT DISTRIBUTED WITHIN THE U.S.

Description of Event or Problem · 1

A PATIENT REPORTED TO BAXTER-(B)(4) THAT ON (B)(6) 2010 THE TRANSFER SET LEAKED. ACCORDING TO THE PATIENT, THE TWIST CLAMP WAS DAMAGED. THE TRANSFER SET WAS REPLACED ON (B)(6) 2010. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP EAST EUROPE SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H09I16081

Patients

Seq Age Sex Outcome Treatment
1