FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED HUMIDIFICATION CHAMBER

MDR report key: 1960464 · Received January 13, 2011

Report

Report Number
9611451-2011-00028
Event Type
Malfunction
Date Received
January 13, 2011
Date of Event
December 9, 2010
Report Date
December 20, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT MR290VX VENTED AUTOFEED HUMIDIFICATION CHAMBERS HAD BEEN DESTROYED AT THE HEALTHCARE FACILITY. WE ARE CURRENTLY IN THE PROCESS OF OBTAINING FURTHER INFORMATION FROM THE HEALTHCARE FACILITY IN ORDER TO ASSIST US WITH THE ANALYSIS AND IDENTIFICATION OF A POSSIBLE ROOT CAUSE OF THE EVENT AS REPORTED. WE WILL PROVIDE A FOLLOW UP REPORT ONCE WE RECEIVE FURTHER INFORMATION AND HAVE COMPLETED OUR ANALYSIS.

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT MR290VX VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE NOT RETURNED TO FPH (B)(4) FOR INVESTIGATION. OUR ANALYSIS IS ACCORDINGLY BASED ONLY ON THE EVENT DESCRIPTION PROVIDED BY THE HEALTHCARE FACILITY. THE HEALTHCARE FACILITY REPORTED THAT THE WATER FEEDSET SPIKES OF THREE MR290VX VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE FOUND LEAKING DURING USE. WITHOUT THE COMPLAINT DEVICE IT IS NOT POSSIBLE TO DETERMINE IF THE LEAK IN THE WATER FEEDSET TUBE CONNECTION IS DUE TO PHYSICAL DAMAGE, INCORRECT ASSEMBLY OR OTHER CAUSES. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. OUR USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." NO PATIENT CONSEQUENCE WAS REPORTED AS A RESULT OF THIS INCIDENT. (B)(4).

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED TO AN FPH FIELD REPRESENTATIVE THAT THE WATER FEEDSET SPIKES OF THREE MR290VX VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE FOUND LEAKING DURING USE. IT WAS ALSO REPORTED THAT THE MR290VX CHAMBERS HAVE BEEN DESTROYED BY THE HEALTHCARE FACILITY. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED TO AN FPH FIELD REPRESENTATIVE THAT THE WATER FEEDSET SPIKES OF THREE MR290VX VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE FOUND LEAKING DURING USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VENTED AUTOFEED HUMIDIFICATION CHAMBER BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED MR290VX 101001

Patients

Seq Age Sex Outcome Treatment
1