FDA Adverse Event Other Summary report: N

ASAHI REXEED SERIES DIALYZERS

MDR report key: 1960219 · Received November 17, 2010

Report

Report Number
8010002-2010-00001
Event Type
Other
Date Received
November 17, 2010
Date of Event
October 14, 2010
Report Date
November 15, 2010
Manufacturer
ASAHI KASEI KURARAY MEDICAL CO., LTD.
Product Code
MSF
PMA / PMN Number
K051187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

WE CHECKED THE QUALITY RECORDS INCLUDING THE MATERIAL INCOMING INSPECTIONS, IN-PROCESS TESTS AND THE PRODUCT RELEASE INSPECTION. HOWEVER, NO ABNORMALITY WAS FOUND. THE LEAKAGE TEST WAS PERFORMED ON EACH PRODUCT DURING THE MANUFACTURING PROCESS. AND NON-CONFORMING PRODUCTS ARE NOT RELEASED FROM THE PLANT. SO THE BROKEN HOLLOW FIBER THAT MAY CAUSE THE BLOOD LEAKAGE COULD BE CAUSE BY INADEQUATE CONDITION OF TRANSPORTATION, STORAGE, AND/OR HANDLING OF PRODUCTS.

Description of Event or Problem · 1

THE PT WAS TREATED ON HEMODIALYSIS USING THE REXEED-21R DIALYZER ON (B)(6) 2010. THE DIALYSIS EQUIPMENT ALARMED THE BLOOD LEAKAGE TWO OR THREE MINUTES AFTER STARTING THE EXTRACORPOREAL CIRCULATION. THE BLOOD CIRCULATION WAS STOPPED IMMEDIATELY. THE DIALYZER (THE PRODUCT FROM THE SAME BOX) AND THE BLOOD TUBE SET WERE REPLACED AND THE DIALYSIS TREATMENT WAS STARTED AGAIN. BUT THE EQUIPMENT WARNED THE BLOOD LEAKAGE AGAIN IN THE SAME TIMING AS THE FIRST TIME. THE DIALYZER (THE PRODUCT FROM THE SAME BOX) AND THE TUBE SET WERE CHANGED TO NEW ONES, AND THE TREATMENT WAS TRIED AGAIN. BUT THE EQUIPMENT DETECTED THE BLOOD LEAKAGE AGAIN. IN THE FOURTH TIME, THE NEW DIALYZER PICKED UP FROM THE DIFFERENT BOX. BUT THE BLOOD LEAK WAS DETECTED JUST AFTER RESTARTING THE BLOOD CIRCULATION. THE PT LOST ABOUT THE 1000ML OF BLOOD BECAUSE THE 250ML OF BLOOD REMAINED IN THE DIALYZER AND BLOOD TUBING SET WAS DISCARDED EACH TIME. THE BLOOD TRANSFUSION WAS NEEDED FOR THE PT IN THE ADDITIONAL HOSPITAL DAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASAHI REXEED SERIES DIALYZERS HEMODIALYZER, RE-USE, HIGH FLUX; MSF MSF ASAHI KASEI KURARAY MEDICAL CO., LTD. REXEED-21R M88N8U

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R BLOOD TRANSFUSION