FDA Adverse Event
Malfunction
Summary report: N
CAREASSIST BED
MDR report key: 1960217
·
Received December 21, 2010
Report
- Report Number
- 9615739-2010-00312
- Event Type
- Malfunction
- Date Received
- December 21, 2010
- Date of Event
- December 1, 2010
- Report Date
- December 1, 2010
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE CASTERS OUT OF SYNC AND THE CASTER SUPPORTS WERE BROKEN. THE TECHNICIAN RE-SYNCED THE CASTERS AND REPLACED THE CASTER SUPPORTS TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE HEAD END BRAKE CASTERS WILL NOT LOCK INTO BRAKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAREASSIST BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM PLUVIGNER | 1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |