FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS

MDR report key: 1960178 · Received December 15, 2010

Report

Report Number
1826988-2010-00840
Event Type
Malfunction
Date Received
December 15, 2010
Date of Event
December 3, 2010
Report Date
December 3, 2010
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K062058
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER CALLED FOR HELP WITH HIS CONTOUR METER. HE PERFORMED CONTROL TESTS WHILE TROUBLESHOOTING AND RECEIVED RESULTS OF 543, 584 AND 590 MG/DL. THE NORMAL CONTROL RANGE WAS 108-149 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS AND NEW METER WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS NBW BAYER HEALTHCARE LLC 7080G 0CC3D11

Patients

Seq Age Sex Outcome Treatment
1 UNK