FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1960139 · Received December 17, 2010

Report

Report Number
1824206-2010-11725
Event Type
Malfunction
Date Received
December 17, 2010
Date of Event
December 7, 2010
Report Date
December 7, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECH FOUND THE HI/LOW HEAD SECTION SLOWLY DRIFTED DOWN. THE HI/LOW HEAD DOWN VALVE WAS STUCK. REPLACED HI/LOW HEAD DOWN VALVE TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

INFO RECEIVED THAT THE HI/LOW HEAD SECTION WOULD NOT STAY UP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1900AD003362

Patients

Seq Age Sex Outcome Treatment
1