FDA Adverse Event
Malfunction
Summary report: N
VERSACARE BED
MDR report key: 1960130
·
Received December 17, 2010
Report
- Report Number
- 1824206-2010-11754
- Event Type
- Malfunction
- Date Received
- December 17, 2010
- Date of Event
- November 30, 2010
- Report Date
- November 30, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECH DISCOVERED, AFTER REMOVING THE COVER FROM THE PAN HOUSING LOGIC BOARD, A BURN ODOR. AFTER FURTHER VISUAL INSPECTION, THE TECH DISCOVERED BURNT COMPONENTS ON THE MOTOR CONTROL BOARD. THE TECH REPLACED THE MOTOR CONTROL BOARD TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
ACCOUNT ALLEGED THAT WHEN RAISING OR LOWERING THE BED, IT WOULD GO INTO TRENDELENBURG. NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VERSACARE BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |