FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 1960130 · Received December 17, 2010

Report

Report Number
1824206-2010-11754
Event Type
Malfunction
Date Received
December 17, 2010
Date of Event
November 30, 2010
Report Date
November 30, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECH DISCOVERED, AFTER REMOVING THE COVER FROM THE PAN HOUSING LOGIC BOARD, A BURN ODOR. AFTER FURTHER VISUAL INSPECTION, THE TECH DISCOVERED BURNT COMPONENTS ON THE MOTOR CONTROL BOARD. THE TECH REPLACED THE MOTOR CONTROL BOARD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

ACCOUNT ALLEGED THAT WHEN RAISING OR LOWERING THE BED, IT WOULD GO INTO TRENDELENBURG. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSACARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment
1