FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 1960129 · Received December 17, 2010

Report

Report Number
1824206-2010-11753
Event Type
Malfunction
Date Received
December 17, 2010
Date of Event
November 30, 2010
Report Date
November 30, 2010
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ACCOUNT SAID THAT THEY ARE NOT GETTING A CLICKING ON THE SIDE COMMUNICATION BOARD. TECH ASKED ACCOUNT TO CHECK PINS 30-31 FOR CONTINUITY ON THE 37 PIN CONNECTOR. NO FURTHER INFO IS AVAILABLE ON THE REPAIR OF THIS BED AT THIS TIME.

Description of Event or Problem · 1

ACCOUNT ALLEGED THAT THE BED EXIT IS NOT ALARMING AT THE BED. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1