FDA Adverse Event
Malfunction
Summary report: N
CAREASSIST BED
MDR report key: 1960129
·
Received December 17, 2010
Report
- Report Number
- 1824206-2010-11753
- Event Type
- Malfunction
- Date Received
- December 17, 2010
- Date of Event
- November 30, 2010
- Report Date
- November 30, 2010
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ACCOUNT SAID THAT THEY ARE NOT GETTING A CLICKING ON THE SIDE COMMUNICATION BOARD. TECH ASKED ACCOUNT TO CHECK PINS 30-31 FOR CONTINUITY ON THE 37 PIN CONNECTOR. NO FURTHER INFO IS AVAILABLE ON THE REPAIR OF THIS BED AT THIS TIME.
Description of Event or Problem · 1
ACCOUNT ALLEGED THAT THE BED EXIT IS NOT ALARMING AT THE BED. NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAREASSIST BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM PLUVIGNER | 1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |