FDA Adverse Event
Malfunction
Summary report: N
REPAIR CORE IMPACTION DRILL
MDR report key: 1960078
·
Received December 17, 2010
Report
- Report Number
- 1811755-2010-02085
- Event Type
- Malfunction
- Date Received
- December 17, 2010
- Date of Event
- November 23, 2010
- Report Date
- November 23, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE HAS BEEN REC'D AT THE MFR FOR INVESTIGATION. AN EVAL WAS CONDUCTED AND THE COMPLAINT WAS CONFIRMED. ACCORDING TO THE INVESTIGATION DETAILS, CORROSION WAS FOUND INSIDE THE DEVICE AND THE BEARING WAS LODGED IN THE SPINDLE HOUSING. THE BEARING, SPINDLE HOUSING, ALONG WITH OTHER COMPONENTS, WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THERE WAS NO REPORTED PT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REPAIR CORE IMPACTION DRILL | BONE CUTTING INSTRUMENT AND ACCESSORIES | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |