FDA Adverse Event Malfunction Summary report: N

INTELLIVUE

MDR report key: 1960054 · Received December 8, 2010

Report

Report Number
1960054
Event Type
Malfunction
Date Received
December 8, 2010
Date of Event
November 16, 2010
Report Date
December 8, 2010
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
MHX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
NC, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

DURING A PROCEDURE, THE SCREEN ON THE PHILIPS INTELLIVUE MP90 WENT BLANK. THE MONITOR WAS REMOVED FROM THE OR AND A SPARE WAS USED TO REPLACE IT. MEDICAL ENGINEERING WAS UNABLE TO DUPLICATE THE PROBLEM, AND PUT THE MONITOR BACK INTO SERVICE. THE MONITOR AGAIN WENT BLANK AND IT WAS AGAIN REMOVED FROM SERVICE. PHILIPS WAS CALLED, AND THE REPRESENTATIVE REPLACED ONE BOARD IN THE CPU (THE BOARDS HAD BEEN REPLACED IN SEPTEMBER). THE MONITOR WAS ALLOWED TO RUN FOR SEVERAL DAYS, AND THEN BEGAN OPERATE INCORRECTLY AGAIN. THE MONITOR WAS RETURNED TO PHILIPS FOR INVESTIGATION. BACK IN SEPTEMBER THE BOARDS IN THE CPU ON THIS UNIT HAD BEEN REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTELLIVUE MONITOR, PHYSIOLOGICAL MHX PHILIPS MEDICAL SYSTEMS MP90 *

Patients

Seq Age Sex Outcome Treatment
1 *