FDA Adverse Event Malfunction Summary report: N

OT ULTRALINK METER

MDR report key: 1960001 · Received January 13, 2011

Report

Report Number
2939301-2011-00464
Event Type
Malfunction
Date Received
January 13, 2011
Report Date
January 1, 2011
Manufacturer
LIFESCAN INC.
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW UP # 1/SUPPLEMENTAL REPORT TEXT 01/25/2011. THE LAY USER/PATIENT'S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS:THE METER INVOLVED IN THIS CASE HAS PASSED TESTING WITH NO FAULTS FOUND. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K) # IS K073231.

Description of Event or Problem · 1

ON (B)(6) 2011, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HIS ONETOUCH ULTRALINK METER WOULD NOT POWER ON. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ON (B)(6) 2010 AT 10PM. THE CCA WAS ADVISED THE PATIENT MANAGES HIS DIABETES WITH HUMALOG INSULIN (AMOUNT NOT SPECIFIED). IT IS NOT KNOWN WHAT ACTION THE PATIENT TOOK AT THE TIME OF THE ALLEGED ISSUE. AN HOUR AFTER THE ALLEGED ISSUE BEGUN, THE PATIENT CLAIMED HE DEVELOPED A "CRAMP." THE PATIENT DENIED RECEIVING MEDICAL TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THERE WAS NO MISUSE OF THE SUBJECT METER. THE CCA TRIED TO WALK THE PATIENT THROUGH RESOLVING THE ALLEGED ISSUE AND DISCOVERED THE METER WOULD NOT POWER ON WHEN PRESSING BUTTONS. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. BASED ON THE INFORMATION PROVIDED, THERE IS NO INDICATION THAT THE REPORTED ISSUE CAUSED OR CONTRIBUTED TO A SERIOUS INJURY SINCE THE PATIENT SYMPTOMS DID NOT CORRELATE WITH LFS'S DEFINITION SUGGESTIVE OF SEVERE HYPOGLYCEMIA OR HYPERGLYCEMIA, NOR RECEIVE MEDICAL INTERVENTION FOR EITHER OF THESE CONDITIONS. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED POWER ISSUE REMAINED UNRESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC. 3059818

Patients

Seq Age Sex Outcome Treatment
1 67 YR