FDA Adverse Event Malfunction Summary report: N

MINIMED QUICK-SET

MDR report key: 19593724 · Received June 24, 2024

Report

Report Number
3003442380-2024-10034
Event Type
Malfunction
Date Received
June 24, 2024
Date of Event
May 22, 2024
Report Date
November 5, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244017450
PMA / PMN Number
K160648
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4 AND MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. PER REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DEVICE HISTORY RECORD (DHR) REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DHR REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD 2426497 THE BATCH 6004011, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DHR REVIEW: THE LOT 6004011 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 77 AND PACKAGED IN THE MACHINE MULTIVAC 08-12 ON 31-OCT-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, ASSEMBLY OF QUICK SET OF THE LOT 3K04923 WAS MANUFACTURED ACCORDING TO THE WI VERSION 26 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 31-OCT-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, ASSEMBLY OF QUICK SET OF THE LOT 3K03773 WAS MANUFACTURED ACCORDING TO THE WI VERSION 26 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 30-OCT-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, ASSEMBLY OF QUICK SET OF THE LOT 3L01725 WAS MANUFACTURED ACCORDING TO THE WI VERSION 26 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 11-NOV-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, TUBE GLUING SET OF THE LOT 3K04908 WAS MANUFACTURED ACCORDING TO THE WI VERSION 40 AND MANUFACTURED IN THE MACHINE 04-05-08, ON 30-OCT-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, TUBE GLUING SET OF THE LOT 3L01400 WAS MANUFACTURED ACCORDING TO THE WI VERSION 41 AND MANUFACTURED IN THE MACHINE 04-05-08, ON 10-NOV-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, TUBE GLUING SET OF THE LOT 3K02798 WAS MANUFACTURED ACCORDING TO THE WI VERSION 40 AND MANUFACTURED IN THE MACHINE 04-05-08, ON 27-OCT-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE; NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: (B)(6).

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT (B)(4). ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. THE INVESTIGATION ASSOCIATED WITH RELATED EVENT (B)(4) HAS BEEN APPROVED AND IS COMPLETE. NO ADDITIONAL ACTION IS REQUIRED, AND THIS COMPLAINT WILL BE CLOSED. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKETING SURVEILLANCE (PMS) PRODUCT TRENDS AND MALFUNCTION ACCORDING TO THE ON MARKET QUALITY REVIEW (OMQR). THE IDENTIFIED MALFUNCTION ADHESIVE PATCH LIFTS OR DETACHES DURING USE, IS NOT ASSOCIATED WITH A DESIGN OR MANUFACTURING-RELATED COMPLAINT ISSUE. THEREFORE, A DETAILED INVESTIGATION OR INSPECTION OF REFERENCE SAMPLES IS NOT REQUIRED. BATCH REVIEW: LOT 6003634 WAS MANUFACTURED ON 14/OCT/2024, IN MACHINE L182, WITH A TOTAL OF (B)(4) PCS. THE BATCH RECORD WAS REVIEWED TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED, AND NO ISSUES WERE FOUND. REVIEW OF THE BATCH RECORD SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCY RELATED TO THIS ISSUE WAS FOUND WITHIN THE DOCUMENTATION. DUE TO THE FOLLOWING BATCH RECORD REVIEW YIELDING NO DISCREPANCIES AND NO NON-CONFORMANCE (NC) WAS GENERATED DURING PRODUCTION. NO FURTHER ACTION IS REQUIRED FOR THIS COMPLAINT. THE ACRYLIC GLUE ON THE ADHESIVE IS NOT APPLIED BY CONVATEC AND THE ADHESIVE MATERIAL HAS BEEN MANUFACTURED AND INSPECTED IN ACCORDANCE WITH ALL SPECIFICATIONS, BATCH DOCUMENTATION AND THE REQUIREMENTS ACCORDING TO 21 CFR PART 820: QUALITY SYSTEM REGULATION AND ISO EN 13485:2016: MEDICAL DEVICES-QUALITY MANAGEMENT SYSTEMS.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN (B)(6). ON 22-MAY-2024, IT WAS REPORTED BY THE PATIENT THAT INFUSION SET WAS LEAKING DUE TO WHICH HYPERGLYCEMIA OCCURS WITHIN ONE DAY OF USE. HIGH BLOOD GLUCOSE LEVEL WAS 363 MG/DL. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
417443 MINIMED QUICK-SET UNO QUICK-SET 60/6 SC1 MECA FPA UNOMEDICAL A/S MMT-399A 6004011 05705244017450

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown