FDA Adverse Event Injury Summary report: N

INTELLIS

MDR report key: 19586974 · Received June 21, 2024

Report

Report Number
3004209178-2024-13320
Event Type
Injury
Date Received
June 21, 2024
Date of Event
June 1, 2024
Report Date
July 10, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LGW
UDI-DI
00643169781702
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID: 3708240, SERIAL#(B)(6), IMPLANTED:(B)(6) 2008, PRODUCT TYPE: EXTENSION. PRODUCT ID: 3998, LOT# V060415, IMPLANTED: (B)(6) 2008, PRODUCT TYPE: LEAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

REP REPORTED THAT LEAD REVISION PLANNED, NO DATE SET UP YET.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM A PATIENT VIA MANUFACTURER REPRESENTATIVE WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). 0-40000 ,1-35440,2-40000,3-40000, HIGH IMPEDANCES WERE REPORTED. LOSS OF STIMULATION AND RETURN OF PAIN WERE REPORTED. TRIED REPROGRAMMING USING UNAFFECTED ELECTRODES. THE ISSUE WAS NOT RESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
218610 INTELLIS STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR LGW MEDTRONIC PUERTO RICO OPERATIONS CO. 97715 00643169781702

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female Required Intervention