FDA Adverse Event Malfunction Summary report: N

R SERIES DEFIBRILLATOR

MDR report key: 19583061 · Received June 21, 2024

Report

Report Number
1220908-2024-02240
Event Type
Malfunction
Date Received
June 21, 2024
Report Date
June 3, 2024
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
UDI-DI
00847946017279
PMA / PMN Number
P160022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION FOR EVALUATION. THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO AN INCORRECT DEVICE SETTING. THE DEVICE IS CONFIGURED TO RUN THE AUTOMATIC TEST USING CPR/MWP/COMPLETE PADZ. LOG REVIEW OBSERVED THAT THE DEVICE WAS CONNECTED TO ONESTEP BASIC PADS.THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
164248 R SERIES DEFIBRILLATOR DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION R SERIES NA 00847946017279

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown