640G INSULIN PUMP MMT-1711K
Report
- Report Number
- 2032227-2024-189664
- Event Type
- Malfunction
- Date Received
- June 21, 2024
- Date of Event
- June 1, 2024
- Report Date
- January 15, 2025
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 000000763000317089
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EN
- Reporter Occupation
- 003
Narratives
PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST AND DISPLACEMENT TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. SUCCESSFULLY UTILIZED CREST AND THUS TO DOWNLOAD HISTORY FILES, TRACES AND COMLINK3 FILES. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM AND PUMP ERROR 35 ALARM CONFIRMED IN THE FORMATTED HISTORY FILE ON (B)(6) 2024 14:35:00.000 AND (B)(6) 2024 14:45:01.000. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND CORROSION ON THE PCBA 1, PCBA 2, FORCE SENSOR AND MOTOR ASSEMBLY NOTED. PLEASE SEE BELOW FOR PUMP ERROR(S)/ALARM(S) NOTED 2 DAYS PRIOR TO THE EVENT DATE 01-JUN-2024 IN THE FORMATTED HISTORY FILE. LOST SENSOR 1 ALERT (780) WAS FOUND ON: (B)(6) 2024 21:31:00.000 AND (B)(6) 2024 12:27:00.000. LOST SENSOR 2 ALERT (781) WAS FOUND ON: (B)(6) 2024 12:43:00.000. UNABLE TO TEST FOR LOST SENSOR ALERT DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. LOST SENSOR ALERT WAS UNKNOWN. INSERT BATTERY ALARM WAS FOUND ON: (B)(6) 2024 12:43:00.000, (B)(6) 2024 14:58:13.000, (B)(6) 2024 16:04:15.000 AND (B)(6) 2024 16:23:11.000. FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS FOUND ON: (B)(6) 2024 14:38:29.000, (B)(6) 2024 14:59:18.000 AND (B)(6) 2024 16:05:24.000. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. INSERT BATTERY ALARM WAS EXPECTED SINCE THE BATTERY WAS REMOVED FROM THE PUMP. UPON CHECKING ON THE POWER DATA/DETAIL TRACE FILE, FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS EXPECTED DUE TO PUMP BATTERY DOES NOT HAVE ENOUGH POWER. THE CUSTOMER HAD USED A NO POWER/DEPLETED BATTERY. UNABLE TO TEST FOR FAILED BATTERY ALERT/BATTERY FAILED ALARM DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS UNKNOWN. PUMP ERROR 130 ALARM WAS FOUND ON: (B)(6) 2024 15:51:17.000 AND (B)(6) 2024 15:51:22.000. PUMP ERROR 130 ALARM WAS CONFIRMED ACCORDING IN THE FORMATTED HISTORY FILE DUE TO CORROSION ON THE MOTOR ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY, A CRACKED BATTERY TUBE THREADS, A SCRATCHED CASE AND A SERIAL NUMBER LABEL MISSING. UNABLE TO PERFORM THE REQUIRED TESTING DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO FATAL ALARM PUMP ERROR 35. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM AND PUMP ERROR 35 ALARM CONFIRMED DUE TO CORROSION ON THE PCBA 1, PCBA 2 AND FORCE SENSOR. ALSO, PUMP ERROR 130 ALARM WAS CONFIRMED ACCORDING IN THE FORMATTED HISTORY FILE DUE TO CORROSION ON THE MOTOR ASSEMBLY NOTED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED OUTSIDE THE UNITED STATES. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED PUMP ERROR 35. THE CUSTOMER REPORTED DISCOMFORT TREATED WITH OTHER. THE EVENT INVOLVED PRODUCT(S) MMT-1711K. TROUBLESHOOTING WAS PERFORMED AND INSTRUCTED THE PUMP WILL BE REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. MMT-1711K WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1197619 | 640G INSULIN PUMP MMT-1711K | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1711K | HG4654H | 000000763000317089 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |