OT ULTRASMART METER
Report
- Report Number
- 2939301-2011-00420
- Event Type
- Injury
- Date Received
- January 11, 2011
- Date of Event
- December 16, 2010
- Report Date
- January 2, 2011
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K) # IS K021819.
ON (B)(6) 2011, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HER ONETOUCH ULTRASMART METER WOULD NOT POWER ON. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ON (B)(6) 2010. THE CCA WAS ADVISED THE PATIENT MANAGES HER DIABETES WITH ACTOS AND METFORMIN PILLS (AMOUNTS NOT SPECIFIED). IT IS NOT KNOWN WHAT ACTION THE PATIENT TOOK AT THE TIME OF THE ALLEGED ISSUE; HOWEVER, STATED SHE EXERCISED. ON (B)(6) 2010, AFTER THE START OF THE ALLEGED ISSUE, THE PATIENT CLAIMED SHE FELT SYMPTOMS OF SHAKY AND THIRSTY. THE PATIENT DENIED RECEIVING MEDICAL TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THERE WAS NO MISUSE OF THE SUBJECT METER. THE CCA TRIED TO WALK THE PATIENT THROUGH RESOLVING THE ALLEGED ISSUE AND DISCOVERED THE METER WOULD NOT POWER ON WHEN PRESSING BUTTONS. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT CLAIMS SHE WAS UNABLE TO TEST HER BLOOD GLUCOSE DUE TO THE REPORTED ISSUE AND REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY AFTER THE ALLEGED METER ISSUE BEGAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT ULTRASMART METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN INC. | 3041404 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Life Threatening |