FDA Adverse Event Malfunction Summary report: N

AUTOFEED HUMIDIFICATION CHAMBER

MDR report key: 1957775 · Received January 11, 2011

Report

Report Number
9611451-2011-00023
Event Type
Malfunction
Date Received
January 11, 2011
Date of Event
December 16, 2010
Report Date
December 17, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE RETURNED AUTOFILL HUMIDIFICATION CHAMBER WAS VISUALLY INSPECTED FOR DAMAGE. RESULTS: A BREAK WAS FOUND IN THE FEEDSET TUBE AT THE POINT WHERE THE TUBE IS INSERTED INTO THE SPIKE. THE SURFACE OF THE BREAK IS ROUGH. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 090802. CONCLUSION: THE SURFACE OF THE SPLIT IN THE FEED SET TUBING WAS ROUGH, SUGGESTING THAT THE SPLIT WAS CAUSED BY PULLING. IT IS POSSIBLE THAT THE SPIKE GOT CAUGHT OR THAT THE TUBING HAD BEEN UNDER TENSION, CAUSING THE TUBE TO BREAK. ALL MR290 CHAMBERS ARE PRESSURE TESTED FOR LEAKS BEFORE THEY LEAVE THE PRODUCTION LINE. THIS SUGGESTS THAT THE FEED SET TUBING WAS SPLIT POST-PRODUCTION. THE INSTRUCTIONS PROVIDED WITH THE MR290 CHAMBERS ADVISE THE USER TO "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT". (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT HUMIDIFICATION CHAMBER IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE CHAMBER AND COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT THE WATERFEED TUBING BROKE OFF OF AN MR290 AUTOFILL HUMIDIFICATION CHAMBER. THE TUBING SEPARATED FROM THE CHAMBER PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT THE WATERFEED TUBING BROKE OFF OF AN MR290 AUTOFILL HUMIDIFICATION CHAMBER. THE TUBING SEPARATED FROM THE CHAMBER PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOFEED HUMIDIFICATION CHAMBER BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED MR290V 090802