FDA Adverse Event Malfunction Summary report: N

BD CONNECTA PLUS3 BLUE

MDR report key: 19568623 · Received June 19, 2024

Report

Report Number
9610847-2024-00172
Event Type
Malfunction
Date Received
June 19, 2024
Date of Event
May 20, 2024
Report Date
June 25, 2024
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMG
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 394602 AND LOT NUMBER 3153598. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD CONNECTA PLUS3 BLUE WAS CRACKED AFTER THE OPERATING ROOM NURSE ESTABLISHED INTRAVENOUS ACCESS FOR THE PATIENT, THE TEE FOR INFUSION WAS CONNECTED TO FACILITATE THE PUSH OF ANESTHETICS, AND ABOUT 2 MINUTES LATER THERE WAS A LEAK AT THE TEE CONNECTION, AND UPON CAREFUL EXAMINATION, A CRACK WAS FOUND IN THE INFUSION TEE

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
672726 BD CONNECTA PLUS3 BLUE STOPCOCK, I.V. SET FMG BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 3153598

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown