FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 19563406 · Received June 18, 2024

Report

Report Number
1645337-2024-07335
Event Type
Injury
Date Received
June 18, 2024
Date of Event
May 1, 2024
Manufacturer
MENTOR TEXAS
Product Code
FTR
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: MATERIAL RUPTURE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON JULY 11, 2024, MENTOR RECEIVED ADDITIONAL INFORMATION INDICATING THAT THE PATIENT WAS A PART OF A CLINICAL STUDY FOR THE IMPLANTED DEVICES. ON JULY 16, 2024, MENTOR ALSO RECEIVED ADDITIONAL INFORMATION INDICATING THAT THE PATIENT ALSO UNDERWENT LEFT BREAST PARTIAL CAPSULECTOMY FOR THE ANTERIOR ASPECT OF THE LEFT CAPSULE WAS THICKENED SIGNIFICANTLY.

Additional Manufacturer Narrative · 0

ON JULY 2, 2024, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON JULY 9, 2024, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SM MPP GEL 350CC BREAST IMPLANT WAS FOUND TO BE RUPTURED AND RECEIVED IN TWO (2) PARTS. ADDITIONALLY, SHELL ABRASION WAS OBSERVED ON THE EDGES OF THE RUPTURE. A MICROSCOPIC EXAMINATION WAS PERFORMED, AND INSTRUMENT DAMAGE WAS NOT FOUND AT THE EDGES OF THE TEAR. THE EVALUATION DETERMINED THAT THE POSSIBLE CAUSE OF THE RUPTURE WAS CONSISTENT WITH NORMAL WEAR. SHELL ABRASION SUGGESTS IN-VIVO FOLDING OR CREASING OF THE DEVICE. THIS MAY BE THE RESULT OF THE FOLLOWING FACTORS: CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE SUCH AS TOO SMALL BREAST POCKET AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. IN SOME CASES, THE BREAST IMPLANTS MAY ALSO WEAR OUT OVER TIME. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY'S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULES, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST RECONSTRUCTION REVISION WITH A 350CC MENTOR MEMORYGEL BREAST IMPLANT ON THE LEFT BREAST. POST-OPERATIVELY, THE PATIENT SUFFERED LEFT BREAST IMPLANT RUPTURE AND LEFT BREAST CAPSULAR CONTRACTURE (BAKER GRADE II). AS A RESULT, THE PATIENT UNDERWENT BILATERAL BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2024. THE REPLACEMENT DEVICES WERE: (LEFT) 350CC MENTOR MEMORYGEL BREAST IMPLANT CATALOG: 3503501BC, LOT: 9981580, SN: (B)(6) AND (RIGHT) 400CC MENTOR MEMORYGEL BREAST IMPLANT CATALOG: 3504001BC, LOT: 9536112, SN: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
794162 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 7630177

Patients

Seq Age Sex Outcome Treatment
1 60 YR Female Required Intervention