FDA Adverse Event Malfunction Summary report: N

TRUSTEEL

MDR report key: 19560432 · Received June 18, 2024

Report

Report Number
3003442380-2024-08814
Event Type
Malfunction
Date Received
June 18, 2024
Date of Event
May 8, 2024
Report Date
November 12, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018488
PMA / PMN Number
K041545
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4. THE INITIAL MDR WITH MANUFACTURING REPORT NUMBER, WAS SUBMITTED ON 19-JUN-2024. UPON COMPLETION OF THE INVESTIGATION, THE MANUFACTURING DATE WAS UPDATED AS 09-AUG-2023. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. PER REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DHR REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DHR REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6001873, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6001873 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 93 AND PACKAGED IN THE MACHINE MULTIVAC 10 ON 09-AUG-2023, WITH A TOTAL OF (B)(4) UNITS. TUBE GLUING THE SUB-ASSEMBLY, TUBE GLUING OF THE LOT 3F03032 WAS MANUFACTURED ACCORDING TO THE WI VERSION 55 AND MANUFACTURED IN THE MACHINE SC05, ON 19-JUN-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, TUBE GLUING OF THE LOT 3F03033 WAS MANUFACTURED ACCORDING TO THE WI VERSION 55 AND MANUFACTURED IN THE MACHINE SC06, ON 18-JUN-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT AN EXTENDED INSPECTION WAS CREATED DUE TO CONTAMINATION, EXTENDED INSPECTION WAS FOUND WITHIN SPECIFICATION CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE; NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. EXTENDED INSPECTION IS NOT RELATED WITH THE CLAIMED MALFUNCTION. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6).

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED INFUSION SET TUBING WAS KNOTTED ON (B)(6) 2024. THE INFUSION SET WAS IN USE FOR 2 HOURS. THE BLOOD GLUCOSE LEVEL AT THE TIME OF FIRST EVENT 380 MG/DL AND PATIENT RESOLVED IT WITH CORRECTION BOLUS VIA PUMP FOLLOWED BY REPLACING INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
492903 TRUSTEEL UNO CONTACT DETACH G29 60/8TCAP 10PK INT FPA UNOMEDICAL A/S 1002835 6001873 05705244018488

Patients

Seq Age Sex Outcome Treatment
1 80 YR Female