FDA Adverse Event Malfunction Summary report: N

EVIS LUCERA ELITE COLONOVIDEOSCOPE

MDR report key: 19560112 · Received June 18, 2024

Report

Report Number
9610595-2024-12250
Event Type
Malfunction
Date Received
June 18, 2024
Report Date
August 23, 2024
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FDF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. NEW INFORMATION ADDED TO THE FOLLOWING FIELDS: A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 5 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. BASED ON THE RESULTS OF THE INVESTIGATION AND THE INFORMATION PROVIDED, IT COULD NOT BE DETERMINED WHAT THE FOREIGN MATERIAL WAS. THERE WAS NO DAMAGE TO THE AREA WHERE THE FOREIGN MATERIAL WAS DETECTED. THEREFORE, THE CAUSE OF THE MATERIAL REMAINING IN THE DEVICE COULD NOT BE DETERMINED. THE LEGAL MANUFACTURE REVIEWED THE CUSTOMER PROVIDED THE CLEANING DISINFECTION AND STERILIZATION (CDS) PROCESSES WHERE NO OBVIOUS DEVIATIONS FROM INSTRUCTIONS FOR USE (IFU) WERE IDENTIFIED. THE EVENT CAN BE DETECTED/PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATE: ¿GIF/CF/PCF-290 SERIES, CHAPTER 3 PREPARATION AND INSPECTION¿ DESCRIBES THE METHODS FOR DETECTION. ¿GIF/CF/PCF-290 SERIES, CHAPTER 5 REPROCESSING THE ENDOSCOPE (AND RELATED REPROCESSING ACCESSORIES)¿ DESCRIBES THE METHODS FOR PREVENTION. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE DEVICE EVALUATION IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE EXHIBITED THE NOZZLE HAVING FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1651916 EVIS LUCERA ELITE COLONOVIDEOSCOPE COLONOVIDEOSCOPE FDF AIZU OLYMPUS CO., LTD. CF-HQ290ZI

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown