FDA Adverse Event Malfunction Summary report: N

TENDRIL ST

MDR report key: 1955838 · Received January 10, 2011

Report

Report Number
2017865-2011-00960
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
October 20, 2009
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
P960013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD EXHIBITED OVERSENSING. THE LEAD POLARITY WAS CHANGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/58 NA

Patients

Seq Age Sex Outcome Treatment
1 76 YR