FDA Adverse Event Malfunction Summary report: N

TENDRIL SDX

MDR report key: 1955829 · Received January 10, 2011

Report

Report Number
2017865-2011-00816
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
September 28, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
P960013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD EXHIBITED LOW IMPEDANCE OF 110 OHMS AND LOW SENSING THRESHOLD BETWEEN 3 AND 0.8 MV. TWO EPISODES OF NOISE WERE ALSO NOTED, WHICH WERE CONSISTENT WITH PRESUMABLE INSULATION DAMAGE. PROGRAMMING WAS CHANGED TO VENTRICULAR ONLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1688TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 68 YR