FDA Adverse Event
Malfunction
Summary report: N
QUICK FLEX LV LEAD
MDR report key: 1955669
·
Received January 10, 2011
Report
- Report Number
- 2017865-2011-00641
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- October 25, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NIK
- PMA / PMN Number
- P030054
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD EXHIBITED HIGH CAPTURE THRESHOLD OF 5.25 V AT 0.5 MS. PROGRAMMING WAS CHANGED TO THE LV RING-RV COIL CONFIGURATION, WHERE CAPTURE THRESHOLD WAS 1.5 V AT 1 MS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | QUICK FLEX LV LEAD | PERMANENT PACEMAKER ELECTRODE | NIK | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1156T/86 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |