FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 1955628 · Received January 10, 2011

Report

Report Number
2017865-2011-00892
Event Type
Injury
Date Received
January 10, 2011
Date of Event
November 17, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

FINAL ANALYSIS FOUND THAT THE PROXIMAL INSULATION WAS ABRADED AT 24 CM FROM THE CONNECTOR PIN, DUE TO FRICTION WITH ANOTHER DEVICE.

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THE ATRIAL LEAD EXHIBITED LOSS OF CAPTURE AND UNDERSENSING. AN X-RAY REVEALED LEAD DISLODGEMENT. THE LEAD WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1888TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention