FDA Adverse Event
Malfunction
Summary report: N
TENDRIL ST
MDR report key: 1955561
·
Received January 10, 2011
Report
- Report Number
- 2017865-2011-00980
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- November 4, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DTB
- PMA / PMN Number
- P960013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. RELIABILITY LABORATORY TECHNICIAN, (B)(4) FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND MEDICAL ADHESIVE ON THE DISTAL RING ELECTRODE. THIS COULD CAUSE THE LEAD TO EXHIBIT NOISE AND HIGH IMPEDANCE.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | DTB | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1888TC/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |