FDA Adverse Event Malfunction Summary report: N

UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN

MDR report key: 1955539 · Received January 10, 2011

Report

Report Number
1423500-2011-00392
Event Type
Malfunction
Date Received
January 10, 2011
Date of Event
December 13, 2010
Report Date
December 13, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION EXPECTED, BUT NOT YET COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF BAXTER'S INVESTIGATION. A 510K NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510K NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS THE SAME OR SIMILAR TO THE PRODUCT DISTRIBUTED WITHIN THE U.S.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A DAMAGED WAS CONFIRMED IN THE LAB. THE RESULTS OF A SAMPLE EVALUATION REVEALED THAT THE TIP OF THE SPIKE WAS BENT INWARD AND BASE OF SPIKE WAS BENT BACKWARDS. A BATCH REVIEW WAS NOT PERFORMED DUE TO UNKNOWN LOT NUMBER. A ROOT CAUSE WAS NOT IDENTIFIED DUE TO INSUFFICIENT INFORMATION. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

A BAXTER SALES REPRESENTATIVE REPORTED A SPIKE WAS BENT WHILE BEING CONNECTED TO THE SOLUTION BAG WITH THE CLEAN FLASH DEVICE. THIS WAS DISCOVERED WHEN THE LID OF THE CLEAN FLASH DEVICE WAS OPENED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1