MEDISENSE OPTIUM
Report
- Report Number
- 2954323-2011-00227
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- December 14, 2010
- Report Date
- January 10, 2011
- Product Code
- NBW
- Removal / Correction Number
- ADC FA1197-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
RETAIN SAMPLES OF PRECISION TEST STRIP LOTS MANUFACTURED WITH HOT-MELT GLUE BATCHES EXHIBITED LOWER THAN EXPECTED READINGS ON CONTINUOUS STORAGE AFTER NINE MONTHS AT 30 DEGREES CELSIUS DURING ROUTINE STABILITY PERFORMANCE TESTING. THIS STABILITY ISSUE COULD LEAD TO THE GENERATION OF INCORRECTLY DEPRESSED BLOOD GLUCOSE RESULTS AND THESE ERRONEOUS RESULTS MAY BE IN THE C, D, OR E ZONES OF PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. (B)(4).
A CUSTOMER REPORTED THE TEST DID NOT START WHEN THE TEST STRIP WAS INSERTED INTO THEIR ADC BLOOD GLUCOSE METER AFTER BLOOD WAS APPLIED. CUSTOMER REPORTED USING AN AFFECTED TEST STRIP LOT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDISENSE OPTIUM | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 46341 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |