SYNCHRON UNICEL DXC 800 PRO CLINICAL SYSTEMS
Report
- Report Number
- 2050012-2011-00096
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- December 16, 2010
- Report Date
- December 16, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
CONTROLS WERE RUN BEFORE THE INCIDENT AND THE RESULTS WERE ACCEPTABLE. A FIELD SERVICE ENGINEER (FSE) FOUND DEPOSIT ON THE REAGENT SYRINGE. FSE REPLACED THE NEW MODULAR CHEMISTRY REAGENT SYRINGE AND THE CREM MODULE. THE FOLLOWING MED-WATCH REPORTS ARE LINKED TO THIS EVENT: 2050012-2011-00107.
CONTROLS WERE RUN BEFORE THE INCIDENT AND THE RESULTS WERE ACCEPTABLE. A FIELD SERVICE ENGINEER (FSE) FOUND DEPOSIT ON THE REAGENT SYRINGE. FSE REPLACED THE NEW MODULAR CHEMISTRY REAGENT SYRINGE AND THE CREM MODULE. THE FOLLOWING MED-WATCH REPORTS ARE LINKED TO THIS EVENT: 2050012-2011-00107. (B)(6).
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUS HIGH CREATININE (CREM) RESULTS FOR TWO PATIENTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE RESULT OF PATIENT ONE WAS REPORTED OUT OF THE LABORATORY. RESULT OF PATIENT TWO WAS NOT REPORTED OUT OF THE LABORATORY. THE SAMPLES WERE REPEATED AND LOWER RESULTS WERE OBTAINED. AMENDED REPORT WAS INITIATED FOR PATIENT NUMBER ONE ONLY. PATIENT NUMBER ONE'S RESULTS WILL BE COVERED UNDER MDR 2050012-2011-00107. TREATMENT WAS NOT INITIATED OR WITHHELD FOR PATIENT NUMBER TWO.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUS HIGH CREATININE (CREM) RESULTS FOR TWO PATIENTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE RESULT OF PATIENT ONE WAS REPORTED OUT OF THE LABORATORY. RESULT OF PATIENT TWO WAS NOT REPORTED OUT OF THE LABORATORY. THE SAMPLES WERE REPEATED AND LOWER RESULTS WERE OBTAINED. AMENDED REPORT WAS INITIATED FOR PATIENT NUMBER ONE ONLY. PATIENT NUMBER ONE'S RESULTS WILL BE COVERED UNDER MDR 2050012-2011-00107. TREATMENT WAS INITIATED OR WITHHELD FOR PATIENT NUMBER TWO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON UNICEL DXC 800 PRO CLINICAL SYSTEMS | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR |