FDA Adverse Event
Injury
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 1955243
·
Received January 10, 2011
Report
- Report Number
- 2017865-2011-00130
- Event Type
- Injury
- Date Received
- January 10, 2011
- Date of Event
- November 22, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT AN ALERT FOR POSSIBLE OUTPUT DAMAGE WAS OBSERVED. AFTER INTERROGATING THE DEVICE THERE WAS ONE VT EPISODE. THE DEVICE RECOGNIZED THE VT, CHARGED AND DELIVERED A SHOCK. LEAD DAMAGE WAS SUSPECTED. HOWEVER, FLUOROSCOPY AND VISUAL INSPECTION REVEALED NO DAMAGE TO THE LEAD. THE LEAD WAS CAPPED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1591/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention | (B)(4) |