FDA Adverse Event Injury Summary report: N

UNIFY CRT-D

MDR report key: 1955232 · Received January 10, 2011

Report

Report Number
2017865-2011-00530
Event Type
Injury
Date Received
January 10, 2011
Date of Event
November 1, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT HIGH IMPEDANCE WAS OBSERVED ON THE RV VECTORS. LEAD IMPEDANCE MEASUREMENTS WERE INTERMITTENTLY OUT OF RANGE. THE POCKET WAS OPENED AND A LOOSE SETSCREW WAS OBSERVED. THE SETSCREW WAS RE-TIGHTENED. THE DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNIFY CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3231-40 NA

Patients

Seq Age Sex Outcome Treatment
1 47 YR Required Intervention (B)(4)