PROXIMATE*PPH PROCEDURE SET
Report
- Report Number
- 3005075853-2011-00088
- Event Type
- Malfunction
- Date Received
- January 10, 2011
- Date of Event
- December 22, 2010
- Report Date
- December 28, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K051301
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
(B)(4). WASHER UNCUT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION WITHOUT STAPLES PRESENT AND WITH THE BREAKAWAY WASHER UNCUT, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT IF THE INDICATOR IS NOT FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE OR IF THE FIRING HANDLE IS NOT FIRMLY SQUEEZED USING STEADY PRESSURE UNTIL THE FIRING HANDLE IS FULLY PARALLEL TO THE INSTRUMENT HANDLE, STAPLES COULD BE DEPLOYED WITHOUT COMPLETELY FORMING AND WITHOUT CUTTING THE WASHER. FOR MORE INFORMATION PLEASE REFER TO THE INSTRUCTIONS FOR USE. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4), AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING A HEMORRHOIDECTOMY PROCEDURE, AT ANASTOMOSIS, THE DEVICE CUT BUT DID NOT STAPLE. THEY HAD TO SEW THE RECTUM BY HAND. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE*PPH PROCEDURE SET | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | G4RJ4C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |