FDA Adverse Event Injury Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 1955150 · Received January 10, 2011

Report

Report Number
2017865-2011-00178
Event Type
Injury
Date Received
January 10, 2011
Date of Event
November 25, 2010
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE DELIVERED INAPPROPRIATE HV THERAPY DUE TO NOISE. REVIEW OF THE STORED EGM SHOWED THE NOISE WAS DUE TO A SUSPECTED LEAD FAILURE ON THE SENSE/PACE PORTION OF THE LEAD. THE LEAD WAS PARTIALLY CAPPED AND REPLACED. THE DEFIB PORTION OF THE LEAD REMAINS ACTIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7001/60 NA

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention